SiPM CAMERA / TECHNICAL PRESENTATION

otSPAD

Temperature-dependent
photon-counting performance

Manuele TuriniExperimental performance · 6 November 2026
Recorded average count rate during the 19.95-minute temperature sweep
otSPAD · Manuele TuriniRecorded thermal sweep01
THE CAMERA

Low-light detection,
with thermal control

A silicon photomultiplier, a dedicated analog front end and embedded pulse counting, operated through otSPAD.

Detector
onsemi MICROFC-30050-SMT-TR
3 × 3 mm active area
Readout
LTC6268 transimpedance amplifier
33 kΩ feedback resistance
Bias
OPA551 bias stage
Temperature-dependent calibration
Cooling
Thermoelectric cooling
AD590 cold-plate temperature sensing
Acquisition
FPGA pulse counting
ADSP-BF518 control and timestamping
Software
Live Scope, data analysis and reporting
Continuous CSV session recording

The sensor combines parallel SPAD microcells into one optical detection channel.

otSPAD · Manuele TuriniHardware configuration used for this measurement02
DETECTOR REFERENCE

MICROFC-30050-SMT-TR

onsemi C-Series silicon photomultiplier · 50 µm microcells

9 mm²

Active area: 3 × 3 mm

2,668

Parallel microcells

35%

Typical PDE at 420 nm

Microcell fill factor72%
Peak sensitivity420 nm
Typical gain6 × 10⁶
Breakdown voltage24.2–24.7 V
Recommended overvoltage1–5 V
Breakdown temperature coefficient+21.5 mV/°C
Dark count rate, typical / max.300 / 860 kcps
Dark current, typical / max.319 / 914 nA
Typical optical crosstalk10%
Typical afterpulsing0.6%
Anode–cathode capacitance920 pF
Microcell recharge time constant159 ns

Manufacturer reference: 21 °C and 2.5 V overvoltage unless otherwise stated. PDE is photon detection efficiency. Dark current is derived from dark count data in the datasheet. onsemi MICROC-SERIES/D, Rev. 9, pp. 1–5.

otSPAD · Manuele TuriniSensor parameters provide the reference for camera measurements03
MEASURED PERFORMANCE

A continuous thermal sweep

Cooling, bias calibration and warm-up recorded in the same acquisition

Measured count rate and cold-plate temperature, with target settings and calibration markers
11,970 consecutive gates100 ms nominal integration19 min 57 s recorded counts

FAST: individual-gate count rate. AVG: live-time-weighted rate over ten consecutive gates. Temperature refers to the cold plate; orange vertical markers show calibration events.

otSPAD · Manuele TuriniSession 082922 · otSPAD 21.10 / BF518 firmware 9.304
TEMPERATURE PERFORMANCE PROFILE

Lower background with cooling

Median count rate falls from 301.3 kcps at plus 19.8 degrees C to 11.1 kcps at minus 22.1 degrees C
27×

Observed background reduction
across the selected conditions

Plate
°C
Median
kcps
Cathode
V
Peltier
% command
+19.8301.329.6620.0
+10.1139.629.47310.4
+5.799.229.37215.6
+1.069.129.27521.4
-4.146.729.17728.4
-13.622.228.97144.0
-22.111.128.79064.0

Selected 29–30 s windows from this run. Temperature and bias settings vary together. Reported values are camera count rates; the sensor datasheet provides the device-level reference conditions.

otSPAD · Manuele TuriniWarm reference: +19.8 °C · Cold reference: −22.1 °C05
OPERATING POINT

Bias follows temperature

Seven recorded calibrations follow an applied temperature coefficient of 21.514 millivolts per degree C
21.5

mV/°C applied compensation slope

The seven recorded calibration points track the sensor’s specified breakdown-voltage temperature coefficient.

Overvoltage setting
2.5 V
Comparator threshold command
2.30 V

The plotted voltage is the cathode command relative to camera ground. The readout uses an approximately +2.5 V virtual-anode reference. Device overvoltage is defined relative to breakdown across the SiPM.

otSPAD · Manuele TuriniCamera commands linked to the detector operating point06
ACQUISITION & SOFTWARE

Counts with their
operating context

Each gate is associated with the camera state in the BF518 before USB transfer.

11,970 / 11,970

Consecutive gate records in this run
No sequence gaps in the acquired data

Per-gate data
Raw counts, measured live time and timestamps
Thermal context
Cold plate, heat sink and ambient temperature
Target, Peltier drive command and fan state
Operating settings
Bias and threshold commands
Calibration and control events
Session record
Continuous CSV files, event history and notes
Portable reports and saved curves
otSPAD · Manuele TuriniLive display and complete session recording07
APPLICATION CONTEXT

Low-light research
on biological samples

A temperature-controlled counting platform for experiments on ultra-weak optical emission.

01

Plant studies

Leaf and plant-surface emission experiments.

02

Skin studies

Optical emission experiments on skin surfaces.

03

Optical collection

Collection optics define the observed area and couple light to the 9 mm² detector.

Temperature, bias and count rate
in one experimental record.

otSPAD · Manuele TuriniotSPAD · SiPM camera platform08
REFERENCE

Measurement & terminology

Data behind this presentation

Measurement: 6 November 2026.
Session: 20261106_082922_854_cd64827e.
Source: samples.csv, events.csv and metadata.json in the supplied report archive.

Background values are medians of the following intervals, in temperature-table order:
59–88 s, 401.5–431.5 s, 476.7–506.7 s, 554.4–584.4 s, 682.1–712.1 s, 795.7–825.7 s, 880.2–910.2 s.

The 27× figure is the ratio between the first and last selected median count rates. All plotted camera curves are derived from the recorded data.

Sensor source: onsemi C-Series datasheet, MICROC-SERIES/D, Rev. 9, February 2022.

Terms used

SiPM
Silicon photomultiplier
SPAD
Single-photon avalanche diode
PDE
Photon detection efficiency
cps / kcps
Counts / thousands of counts per second
TIA
Transimpedance amplifier
FPGA
Field-programmable gate array
CSV
Comma-separated values
Overvoltage
Reverse bias above device breakdown