Executive summary
The completed measurement shows a low-noise, stable dark-count stream after an initial thermal settling period. The full ten-minute run has 8,446 cps average counts and F = 1.329; using the final five minutes selected in advance of the sensitivity calculations yields 8,420 cps, a measured standard deviation of 91.8 counts per 1 s and F = 1.001. This closely matches a Poisson count process at the one-second time scale. An independent acquisition with 60 direct 10-second gates supports the same Poisson-compatible interpretation, within the larger variance uncertainty of the longer-gate sample.
| Recorded setup | Value | Recorded setup | Value |
|---|---|---|---|
| Detector | MICROFC-30050-SMT-TR | Gate / run | 1 s / 600 s |
| Active sensor area | 3 × 3 mm² | SiPM discriminator | TLV3501, 10 kΩ feedback |
| Cold-plate setpoint | −25.0 °C | Measured plate, first → last | −24.617 → −25.017 °C |
| Bias command | 28.727 V | Overvoltage setting | 2.5 V (commanded) |
| Comparator threshold | 2.300 V | Cooling command, first → last | 73.5 → 76.4% |
| Acquisition clock | 80 MHz | Gate records / losses | 600 / 0 |
1. Direct observations MEASURED
Warm-up and baseline selection
The initial count rate decreases as the cold plate approaches its final reading and the TEC controller settles. To avoid conflating this drift with the intrinsic counting noise, only gate sequences 301–600 (300 full seconds) are used for the baseline sensitivity calculations.
The selected interval has mean 8,420 cps, 95% approximate confidence interval 8,409.5–8,430.4 cps for its mean under weak serial dependence, and a residual fitted trend of -0.093 counts/s² (linear-trend p = 0.128). The residual trend is not statistically significant at 5%; its possible systematic impact is not guaranteed negligible in later runs.
Statistical interpretation
At a one-second gate, the expected Poisson standard deviation is 91.8 counts. The measured value is 91.8. The Fano factor, Var(N)/E(N), is 1.001 with an approximate 95% interval of 0.858–1.183.
Lag-1 autocorrelation is +0.018, and no selected gate exceeds ±4 sample standard deviations. This supports—but does not prove—a near-Poisson counting model on this observation scale.
Effect of longer integrations
Adjacent one-second gates were added without changing their order. The resulting variance can be compared with the Poisson expectation. Long integrations contain far fewer independent blocks and cannot be characterized to the same confidence from this ten-minute file.
| Integration | Independent blocks | Observed Fano | Reliability |
|---|---|---|---|
| 1 | 300 | 1.001 | Exploratory |
| 2 | 150 | 0.987 | Exploratory |
| 5 | 60 | 0.970 | Exploratory |
| 10 | 30 | 0.701 | Exploratory |
| 20 | 15 | 0.469 | Exploratory |
| 30 | 10 | 0.423 | Exploratory |
| 60 | 5 | 0.643 | Limited — few samples |
| 120 | 2 | 0.948 | Limited — few samples |
Independent 10-second gate validation MEASURED
A second independent 600-second dark-count acquisition was conducted at the same −25 °C cold-plate setpoint, now using 60 direct 10-second gates. The new run started after the original one-second measurement and therefore did not exhibit its initial thermal settling. All 60 gate records were continuous, with no reported missing gates or FIFO overflows.
| Metric | Original 1 s run | Independent 10 s run |
|---|---|---|
| Samples / nominal run time | 600 × 1 s / 600 s | 60 × 10 s / 600 s |
| Full-run mean dark count rate | 8,445.7 cps | 8,436.1 cps |
| Mean rate, final 300 s | 8,420.0 cps | 8,428.8 cps |
| Final-window per-gate standard deviation | 91.8 counts / 1 s | 234.2 counts / 10 s |
| Final-window Fano factor | 1.001 (300 samples) | 0.651 (30 samples) |
| Approx. 95% Fano interval | 0.858–1.183 | 0.413–1.176 |
| Cold-plate reading at first / last gate | -24.617 / -25.017 °C | -25.017 / -25.017 °C |
| Missing gates / FIFO overflow | 0 / 0 | 0 / 0 |
Comparison with the Poisson expectation
Using the original stable baseline of 8,420.0 counts/s, the ideal prediction for a 10-second gate is 84,200 counts with a standard deviation of 290.2 counts. The new stable 10-second gates contain 84,288.3 counts on average, with sample standard deviation 234.2 counts. The stable mean-rate difference between runs is 8.9 counts/s (0.105%).
Independent run source: 25_10s.zip, session 20261109_153010_825_017fa285. The temperature sensor monitors the cold plate; SiPM junction temperature is not directly measured. The slightly different baseline rates need not reflect a change in photon detection efficiency.
2. Estimated optical detection threshold MODEL, NOT MEASURED
For a known stationary dark background of D = 8,420 counts/s, let t be the integration interval, p the detector PDE, and z the chosen ratio S/N. Defining S/N as the expected net signal counts divided by the standard deviation of dark counts, the idealized minimum photon flux incident on the active area is:
Background counts: B = D × t
Ideal Poisson background noise: σB = √(D × t)
Required detected signal: S = z × σB
Incident photon flux: Φmin = z × √(D × t) / (p × t)
= z × √D / (p × √t)
These are thresholds in photons arriving at the active surface per second for a source at the specified wavelength, not the number of photons that can be identified one by one. The 35% PDE is the onsemi datasheet's nominal 420 nm, 21 °C, +2.5 V overvoltage value; its exact value at the cold junction and in this comparator chain was not calibrated. At S/N=1 the signal is at the noise level—not a reliable detection declaration. S/N=3 or 5 is an amplitude criterion, not a guaranteed detection probability or formal detection limit.
Photon flux matrix — S/N = 1, 3 and 5
Values in incident photons per second, across integrations. Defaults: 35% assumed PDE, known Poisson background. Changing PDE or the background treatment recomputes the table in your browser.
| Integration | S/N = 1 | S/N = 3 | S/N = 5 |
|---|---|---|---|
| 1 s | 262.2 | 786.5 | 1,310.9 |
| 2 s | 185.4 | 556.2 | 926.9 |
| 5 s | 117.2 | 351.7 | 586.2 |
| 10 s | 82.9 | 248.7 | 414.5 |
| 20 s | 58.6 | 175.9 | 293.1 |
| 30 s | 47.9 | 143.6 | 239.3 |
| 60 s | 33.8 | 101.5 | 169.2 |
| 120 s | 23.9 | 71.8 | 119.7 |
Absolute signal counts: the corresponding mean number of additional detected photoevents is simply S = p × Φmin × t. Long integration can reveal a weaker average flux, not eliminate dark events or establish that a particular individual count was optical.
3. Alternative C-Series sensors SIMULATED — NOT TESTED
onsemi specifies typical dark count rates at 21 °C and 2.5 V overvoltage of approximately 300 kcps for the 3 × 3 mm² C-Series and 30 kcps for the 1 × 1 mm² variants. For an illustrative projection to the present cold-plate setup, each candidate is scaled by its nominal room-temperature DCR ratio relative to the tested MICROFC-30050:
Dcandidate(−25 °C plate, simulated)
= Dmeasured,30050 × DCRnominal,candidate(21 °C)
/ DCRnominal,30050(21 °C)
Every candidate inherits the same relative temperature reduction, threshold acceptance, front-end counting behavior and 2.5 V overvoltage as the measured unit. None of those assumptions was experimentally validated for the other parts. PDE values refer to the nominal spectral peak at 420 nm under datasheet reference conditions, not measured cold PDE.
| onsemi SiPM | Area mm² | PDE at 420 nm | Nominal DCR, 21 °C (kcps) | Projected DCR at plate −25 °C (cps) | S/N=3, 1 s (photons/s) | Flux density (photons/s/mm²) |
|---|---|---|---|---|---|---|
| MICROFC-30050 TESTED | 9 | 35.0% | 300 | 8,420.0 | 786.5 | 87.4 |
| MICROFC-30035 SIMULATED | 9 | 31.0% | 300 | 8,420.0 | 888.0 | 98.7 |
| MICROFC-30020 SIMULATED | 9 | 24.0% | 300 | 8,420.0 | 1,147.0 | 127.4 |
| MICROFC-10035 SIMULATED | 1 | 31.0% | 30 | 842.0 | 280.8 | 280.8 |
| MICROFC-10020 SIMULATED | 1 | 24.0% | 30 | 842.0 | 362.7 | 362.7 |
| MICROFC-10010 SIMULATED | 1 | 14.000000000000002% | 30 | 842.0 | 621.8 | 621.8 |
4. Measurement limits and reproducibility
What has been demonstrated
Two complete, independent 600-second dark-count acquisitions (1 s and 10 s gates), measured gate timing, temperature telemetry and stable final-window counting variance consistent with Poisson statistics. No missing gate sequence and no logged FIFO overflow. The stabilized measurement reflects the complete SiPM, TIA, comparator, bias and FPGA counting chain.
What remains unmeasured
Actual SiPM junction temperature; absolute PDE at −25 °C; calibrated photon flux at the sensor; the transient response to a weak illuminated source; true dark leakage current in amperes; dark count spectral characteristics on sub-second scales; and the minimum signal with explicitly verified false-alarm and detection probabilities.
Important qualification about temperature
The temperature measurement is on the cold side of the thermoelectric assembly. The SiPM is thermally coupled to this surface through a 0.8 mm printed circuit board. Thickness alone does not determine its thermal resistance: the copper layers, via density, contact pressure, interface material and heat load all matter. Therefore the label “−25 °C” throughout this report refers to the cold-plate setpoint and sensor measurement, not a verified semiconductor junction temperature.
Statistical and methodological choices
- Analysis data: all 600 raw gate counts retained; no outliers removed from the primary run.
- Primary noise window: last 300 consecutive 1 s gates, sequences 301–600, selected to reduce early thermal drift.
- Mean DCR = arithmetic mean of counts per observed one-second gate; standard deviation uses n−1 denominator.
- Fano = sample variance / sample mean; approximate interval assumes weak dependence and stationary counts.
- Re-binning: sum non-overlapping consecutive blocks; block Fano is descriptive at large t due to small sample size.
- Independent direct 10-second validation: 60 gates of 10 s, last 30 compared with the last 300 one-second gates and their non-overlapping 10-second sums; the larger sample uncertainty is retained in interpretation.
- Sensitivity estimates use the observed DCR and ideal Poisson noise, assuming one detection per detected photon and no additional source excess noise.
- Independent background calibration modes propagate Poisson errors but do not model baseline drift or nuisance parameters.
- Simulated device data extrapolate room-temperature nominal DCR ratios, not manufacturer-certified cold performance.
Primary references
- onsemi, C-Series SiPM Sensors, MICROC-SERIES/D, Rev. 9 (February 2022) — active areas, nominal 21 °C DCR, peak PDE at 420 nm, overvoltage and temperature coefficients.
- Currie, L. A., Detection and quantification limits: basic concepts (2004) — distinction between critical level, detection limit and quantification limit.
- Officina Turini, SPAD/SiPM prototype — project description and instrument architecture.
- Experimental source:
25_1s.zip(samples.csv,events.csv,metadata.json), session20261109_151913_128_5c72b3e9, started at2026-11-09T15:19:13.128Z. - Independent validation:
25_10s.zip(samples.csv,events.csv,metadata.json) with direct 10 s gates.