Can a cooled SiPM reveal a signal long after the light is gone?
otSPAD was developed to measure extremely weak optical signals using a programmable silicon photomultiplier, active temperature control, analog signal conditioning, a fast comparator, an FPGA counter and a DSP-based acquisition system. This experiment tests the complete measurement chain under deliberately simple laboratory conditions.
A small, yellow-green phosphorescent coating associated with an intact Am-241-bearing specimen was observed through a 55 mm focal-length objective in a light-tight enclosure. An earlier one-hour run recorded a strong, slowly decreasing count rate. More than seven hours after that run ended, a new measurement was made without re-exciting or repositioning the specimen prior to its first phase.
Record a stable count rate for about 1,001 seconds with the sample in the optical setup.
Pause counting, turn off SiPM bias and remove the specimen from the enclosure.
Recalibrate and allow the previously characterized startup transient to settle before resuming.
Record a further 867 seconds with the specimen absent and compare stabilized sections.
The signal drops back to the instrument baseline.
Across the first phase, the event rate remained close to 13.7 kcps. During the deliberate pause, the specimen was removed and the detector was recalibrated. The resumed measurement was around 11.6 kcps — the previously characterized dark count range of this cooled setup.
| Measure | Specimen present | After specimen removal |
|---|---|---|
| Recorded gates | 10,010 | 8,673 |
| Recorded measurement duration | 1,001.0 s | 867.3 s |
| Mean count rate · final 300 s of each phase | 13,708.0 cps | 11,614.6 cps |
| Fano factor · final 300 s (raw 100 ms gates) | 0.955 | 0.998 |
| Cold-plate temperature · stabilized mean | ~−19.95 °C | ~−19.99 °C |
| Commanded bias | 28.822 V | 28.819 V |
| Comparator threshold | 2.300 V | 2.300 V |
Stabilized means use the last 300 seconds of each phase, to avoid overweighting an early post-calibration settling component. Temperature refers to the cold plate, not a direct junction-temperature measurement.
“The important result is not that the counter detected a bright source. It is that a small, repeatable count-rate excess remained measurable after hours — and disappeared when the specimen was taken away.”
From afterglow to a controlled removal test.
The earlier acquisition began near 68 kcps and decayed over an hour, reaching approximately 21.9 kcps in its final five minutes. The following session started more than seven hours later at approximately 13.7 kcps with the specimen still present. This is consistent with a very slowly fading optical contribution, but it is not by itself a measurement of the radiative excitation mechanism.
The difference is large compared with counting noise.
Using the final five minutes from each phase, the excess rate is 2,093.4 cps, corresponding to 18.0% of the measured dark baseline. The Poisson-scale fluctuations of 100 ms gates are only a few tens of counts, much smaller than the mean separation between the two populations.
Using non-overlapping 10-second blocks, the within-session statistical 95% confidence interval for the mean excess is approximately 2,079–2,108 cps. This narrow interval quantifies counting statistics and short-term variation; it does not account for systematic uncertainty introduced by calibration, optical alignment or changes to the specimen.
A real instrument — not a simulated detector.
The measurement used the Officina Turini otSPAD prototype, based on an onsemi C-Series SiPM and a programmable analog front end. The comparator had already been stabilized with positive feedback following earlier high-rate testing.
The counter is implemented in an FPGA and read through a Blackfin DSP acquisition system. This experiment recorded 18,683 individual gate values with an approximately 100 ms integration time, preserving time ordering and monitoring hardware state throughout the session.
Operating conditions
SiPM: MICROFC-30050-SMT-TR (3 × 3 mm²) · TLV3501 comparator with 10 kΩ feedback · threshold 2.300 V · programmed overvoltage 2.5 V · cold plate ≈ −20 °C · gate 100 ms · 55 mm objective · light-tight improvised enclosure.
Strong evidence of a specimen-associated signal. No exaggerated claims.
What was demonstrated
The prototype reliably discriminated two different count-rate levels in one instrument session: one with the specimen present and one after it was removed. The difference survived for more than seven hours after the earlier run, and the detector returned close to its dark baseline on removal.
What remains to be isolated
The test does not uniquely identify the cause of the excess. A stored afterglow, continuous radiation-driven emission, direct radiation interactions in the SiPM, residual ambient light, and any calibration-related systematic effects cannot be fully separated without further controls.
In particular, the same intervention changed both specimen presence and the detector calibration state. A matched nonradioactive phosphor control or an optically opaque control with the specimen otherwise unchanged would be needed to specifically establish radiation-induced luminescence. The experiment also provides no calibrated photon flux, scintillation yield or source activity measurement.
Why it matters for otSPAD
A detector designed for very low photon flux should deliver more than impressive specifications. It should reveal a real, faint and persistent signal — and then return to a reproducible reference. This experiment is a practical demonstration of the prototype's sensitivity, electronic stability and ability to make a meaningful differential measurement.
Explore the otSPAD prototype and its technical documentationReproducible numbers, explicit limits.
Source files: longRun.zip (earlier one-hour observation) and finalTest.zip (subsequent controlled removal measurement). For the principal before/after comparison, rates were computed from the raw gate counts and recorded live-gate ticks; descriptive statistics use the final 300 s of each phase. Recorded UTC timestamps and events locate the pause and calibration. No values are fabricated or extrapolated across the pause.
- Officina Turini — Programmable system for single-photon detection, prototype overview.
- onsemi — C-Series SiPM datasheet, MICROFC-30050 family specifications.
- ATSDR / NCBI — Americium-241 decay scheme, radionuclide background; not a measurement of this specimen.
Experimental note, 10 November 2026. All plotted data are from the prototype. Statistical reference curves are calculated from measured means and explicitly labeled as references.